Your browser doesn't support javascript.
loading
Show: 20 | 50 | 100
Results 1 - 2 de 2
Filter
Add filters








Year range
1.
Rev. bras. anestesiol ; 70(4): 388-397, July-Aug. 2020. tab, graf
Article in English, Portuguese | LILACS | ID: biblio-1137193

ABSTRACT

Abstract Background and objectives: The measurement of hemoglobin concentration (Hb) by co-oximetry is an innovative technique that offers efficiency and agility in the processing of information regarding the measurement of Hb obtained through continuous, non-invasive and rapid monitoring. Because of this attribute, it avoids unnecessary exposures of the patient to invasive procedures by allowing a reduction in the number of blood samples for evaluation and other unnecessary therapies. It also helps to make decisions about the need for transfusion and how to handle it. The objective of this study is to compare the performance offered to obtain Hb values between the Masimo Corporation (Irvine, CA, USA) instrument and the standard gold tool (laboratory examination). Contents: The study corresponds to a systematic review followed by meta-analysis, which included fully registered full-text clinical trials published from 1990 to 2018. PubMed, Cochrane, Medline, Embase and Web of Science databases were investigated. The mean overall difference found between the non-invasive and invasive methods of hemoglobin monitoring was 0.23 (95% CI -0.16, 0.62), that is, it did not present statistical significance (p = 0.250). The results of the analysis of heterogeneity within and between the studies indicated high levels of inconsistency (Q = 461.63, p < 0.0001, I2 = 98%), method for Hb values. Conclusions: Although the mean difference between noninvasive measurements of Hb and the gold standard method is small, the co-oximeter can be used as a non-invasive "trend" monitor in detecting unexpected responses at Hb levels.


Resumo Justificativa: A medida da concentração de hemoglobina (Hb)por co-oximetria é uma técnica inovadora que oferece eficiência e agilidade no processamento das informações referentes à medida da concentração de hemoglobina obtida por meio de monitorização contínua, não-invasiva e rápida. Por conta desse atributo, evita exposições desnecessárias do paciente a procedimentos invasivos ao possibilitar redução da quantidade de amostras sanguíneas para avaliação e de outras terapêuticas desnecessárias. Além disso, auxilia a tomada de decisões quanto à necessidade de transfusão e quanto ao manejo da mesma. Objetivo: Comparar o desempenho oferecido para a obtenção dos valores de concentração de hemoglobina entre medida não invasiva da Hb e a ferramenta padrão ouro (exame laboratorial). Conteúdo: O estudo corresponde a uma revisão sistemática seguida de metanálise que incluiu ensaios clínicos devidamente registrados com texto completo, publicados a partir de 1990 até 2018. Foram investigadas as bases de dados PubMed, Cochrane, Medline, Embase e Web Of Science. A diferença média global encontrada entre os métodos não invasivo e invasivo de monitorização da hemoglobina foi de 0,23 (95% IC -0,16; 0,62), ou seja, não apresentou significância estatística (p = 0,250). Os resultados da análise de heterogeneidade dentro e entre os estudos, apontou níveis elevados de inconsistência (Q = 461,63, p< 0,0001, I2 = 98%). Conclusão: Embora a diferença média entre as medidas não invasivas da Hb e o método padrão ouro sejam pequenas, o co-oxímetro pode ser utilizado como um monitor não invasivo de "tendência" na detecção de alterações inesperadas nos níveis de Hb.


Subject(s)
Humans , Hemoglobins/analysis , Oximetry/methods , Monitoring, Physiologic/methods , Research Design , Blood Transfusion , Clinical Trials as Topic/methods
2.
J. bras. pneumol ; 40(3): 222-228, May-Jun/2014. tab, graf
Article in English | LILACS | ID: lil-714687

ABSTRACT

Objective: To evaluate the behavior of oxygen saturation curves throughout the six-minute walk test (6MWT) in patients with COPD. Methods: We included 85 patients, all of whom underwent spirometry and were classified as having moderate COPD (modCOPD, n = 30) or severe COPD (sevCOPD, n = 55). All of the patients performed a 6MWT, in a 27-m corridor with continuous SpO2 and HR monitoring by telemetry. We studied the SpO2 curves in order to determine the time to a 4% decrease in SpO2, the time to the minimum SpO2 (Tmin), and the post-6MWT time to return to the initial SpO2, the last designated recovery time (RT). For each of those curves, we calculated the slope. Results: The mean age in the modCOPD and sevCOPD groups was 66 ± 10 years and 62 ± 11 years, respectively. At baseline, SpO2 was > 94% in all of the patients; none received supplemental oxygen during the 6MWT; and none of the tests were interrupted. The six-minute walk distance did not differ significantly between the groups. The SpO2 values were lowest in the sevCOPD group. There was no difference between the groups regarding RT. In 71% and 63% of the sevCOPD and modCOPD group patients, respectively, a ≥ 4% decrease in SpO2 occurred within the first minute. We found that FEV1% correlated significantly with the ΔSpO2 (r = −0.398; p < 0.001), Tmin (r = −0.449; p < 0.001), and minimum SpO2 (r = 0.356; p < 0.005). Conclusions: In the sevCOPD group, in comparison with the modCOPD group, SpO2 was lower and the Tmin was greater, suggesting a worse prognosis in the former. .


Objetivo: Avaliar o comportamento da curva de saturação de oxigênio durante o teste de caminhada de seis minutos (TC6) em pacientes com DPOC. Métodos: Incluímos 85 pacientes e todos realizaram espirometria, sendo classificados como portadores de DPOC moderada (DPOCm, n = 30) ou grave (DPOCg, n = 55). Todos os pacientes realizaram TC6 em um corredor de 27 m com monitoramento contínuo da SpO2 e FC por telemetria. A partir das curvas de SpO2, foram analisados os tempos para atingir a queda de 4% da SpO2, para atingir a SpO2 mínima (Tmin) e para a recuperação da SpO2 após o TC6 (TR). Foram calculadas as inclinações dessas curvas. Resultados: A média de idade nos grupos DPOCm e DPOCg foi de 62 ± 11 anos e 66 ± 10 anos, respectivamente. Todos os pacientes iniciaram o teste com SpO2 > 94%, nenhum recebeu suplementação de oxigênio durante o TC6, e não houve interrupções. A distância percorrida no TC6 não apresentou diferença significativa entre os grupos. Os menores valores da SpO2 ocorreram no grupo DPOCg. Não houve diferença no TR entre os grupos, e 71% e 63% dos pacientes nos grupos DPOCg e DPOCm, respectivamente, apresentaram queda de SpO2 ≥ 4% até o primeiro minuto. O VEF1% apresentou correlações significativas com ΔSpO2 (r = −0,398; p < 0,001), Tmin (r = −0,449; p < 0,001) e SpO2 mínima (r = 0,356; p < 0,005). Conclusões: As curvas dos pacientes do grupo DPOCg em relação às do grupo DPOCm apresentaram valores menores de SpO2 e maior Tmin, sugerindo um pior prognóstico nos primeiros. .


Subject(s)
Aged , Female , Humans , Male , Middle Aged , Oxygen Consumption , Pulmonary Disease, Chronic Obstructive/physiopathology , Walking/physiology , Exercise Test , Exercise Tolerance , Maximal Expiratory Flow-Volume Curves , Oxygen/metabolism , Pulmonary Disease, Chronic Obstructive/blood , Pulmonary Disease, Chronic Obstructive/classification , Spirometry